Scanning Electron Microscopy

Allen R. Sampson
Advanced Research Systems
December 2, 1996


Index

Section 1 - Introduction

1.0 Introduction to scanning electron microscopy
1.1 A brief history of scanning electron microscopy

Section 2 - Electron optics

2.0 Electron beam generation
2.01 Tungsten filament cathode
2.02 LaB6 cathode
2.03 Field emission gun
2.04 Cathode comparison

2.1 Electron beam lenses and apertures
2.2 Electron beam deflection
2.3 Electron beam aberrations

Section 3 - Electron beam / specimen interactions

3.0 Secondary electron emissions
3.1 Transmitted electrons
3.2 Induced specimen current
3.3 Backscatter electron emission
3.4 Auger electron emission
3.5 X-ray emission

Section 4 - Image presentation and interpretation

4.0 Analog image display and recording
4.1 Digital image display and recording
4.2 Topographical imaging
4.2.0 Secondary imaging
4.2.1 Differential imaging

4.3 Compositional imaging
4.3.0 Backscatter electron emission
4.3.1 Auger electron emission
4.3.2 X-ray emission

4.4 Structural imaging
4.4.0 Induced speciment current
4.4.1 Selected area electron channeling

4.5 Property imaging
4.5.0 Voltage contrast
4.5.1 Electron beam induced current (EBIC)
4.5.2 Magnetic contrast imaging

Section 5 - Elemental analysis

5.0 Backscatter electron emission
5.1 Auger electron emission
5.2 X-ray emission

Section 6 - Sample preparation

6.0 General requirements


1.0 Introduction to scanning electron microscopy

Electron microscopy takes advantage of the wave nature of rapidly moving electrons. Where visible light has wavelengths from 4,000 to 7,000 Angstroms, electrons accelerated to 10,000 KeV have a wavelength of 0.12 Angstroms. Optical microscopes have their resolution limited by the diffraction of light to about 1000 diameters magnification. Electron microscopes, so far, are limited to magnifications of around 1,000,000 diameters, primarily because of spherical and chromatic aberrations. Scanning electron microscope resolutions are currently limited to around 25 Angstroms, though, for a variety of reasons.

The scanning electron microscope generates a beam of electrons in a vacuum. That beam is collimated by electromagnetic condensor lenses, focussed by an objective lens, and scanned across the surface of the sample by electromagnetic deflection coils. The primary imaging method is by collecting secondary electrons that are released by the sample. The secondary electrons are detected by a scintillation material that produces flashes of light from the electrons. The light flashes are then detected and amplified by a photomultiplier tube.

By correlating the sample scan position with the resulting signal, an image can be formed that is strikingly similar to what would be seen through an optical microscope. The illumination and shadowing show a quite natural looking surface topography.

There are other imaging modes available in the SEM. Specimen current imaging using the intensity of the electrical current induced in the specimen by the illuminating electron beam to produce an image. It can often be used to show subsurface defects. Backscatter imaging uses high energy electrons that emerge nearly 180 degrees from the illuminating beam direction. The backscatter electron yield is a function of the average atomic number of each point on the sample, and thus can give compositional information.

Scanning electron microscopes are often coupled with x-ray analysers. The energetic electron beam - sample interactions generate x-rays that are characteristic of the elements present in the sample. Many other imaging modes are available that provide specialized information.

1.1 A brief history of scanning electron microscopy

The first, true scanning electron microscope (SEM) was developed and described in 1942 by Zworykin et al.1. The instrument described eventually consisted of an inverted column (electron gun at the bottom), three electrostatic lenses and electromagnetic scan coils placed between the second and third lenses. A photomultiplier tube detected the scintillations on a phosphor screen caused by the secondary electron emissions. A previous instrument utilizing a scanning electron beam and described by M. von Ardenne2,3 in 1938 was actually the first scanning transmission electron microscope (STEM), using the electrons passing through a thin sample for imaging.

At Cambridge University in 1948, C. W. Oatley began construction of an SEM based on Zworykin's. Graduate student D. McMullan4 described this work in a doctoral dissertation where they had claimed a resolution of 500 Angstroms. Further work, reported by K. C. A. Smith5, made large changes to the electron optics. The electrostatic lenses were replaced with electromagnetic coils, a double deflection scanning system was added as were stigmator coils. Additionally, Smith utilized non-linear signal processing (gamma processing) to help improve imaging.

The next major improvements were made to the signal collection process. The original phosphor screen / photomultiplier used by Zworykin was improved by the addition of a light pipe. The light pipe allowed for direct optical coupling between the scintillator and the photomultiplier tube, greatly improving efficiency6. This was accomplished by T. E. Everhart and R. F. M. Thornley, and the detector arrangement carries their names as the Everhart-Thornley detector.

All of these improvements were combined in one instrument. Once again at Cambridge University, R. F. W. Pease and W. C. Nixon created the SEM V, utilizing the inverted column, electromagnetic lenses, double deflection scan system, stigmation coils and the Everhart-Thornley detector. This instrument became the basis for the first commercial SEM, the Cambridge Scientific Instruments Mark I, first available in 1965.


References, Section 1

1. V. K. Zworykin, J. Hiller and R. L. Snyder, ASTM Bull. 117, 15 (1942)

2. M. von Ardenne, Z. Phys., 109, 553 (1938)

3. M. von Ardenne, Z. Techn. Phys., 19, 407 (1938)

4. D. McMullen, Ph.D. Dissertation, Cambridge University (1952)

5. K. C. A. Smith, Ph.D. Dissertation, Cambridge University (1956)

6. T. E. Everhart and R. F. M. Thornley, J. Sci. Instr., 37 246 (1960)


2.0 Electron optics

2.0 Electron beam generation

The electron gun in a scanning electron microscope is the source for the electron beam used to probe the sample. Electrons are emitted from a cathode, accelerated by passage through electrical fields and focussed to a first optical image of the source. The size and shape of the apparent source, beam acceleration and current are the primary determining factors in the performance and resolution of a scanning electron microscope.

2.01 Tungsten filament cathode

Fig. 1 - typical SEM electron gun

Figure 1 depicts a typical electron gun for an SEM. A bent tungsten wire filament, with a diameter of around 100 micrometers, is spot welded to metal posts. These posts are embedded in a ceramic holder and extend out the other side to provide electrical connections. In operation, the filament will be heated by passing an electrical current through it. Optimum filament temperature for the thermionic emission of electrons is around 2700 degrees Kelvin1.

The accelerating voltage, generally between -500 Volts and -50,000 Volts DC, is applied to the Wehnelt cylinder. Resistive self-biasing is usually used where an adjustable bias resistance connects the filament to the accelerating voltage. The biasing brings the filament slightly more positive than the Wehnelt. The anode is connected to electrical ground.

Without the Wehnelt and anode, electrons emitted from the filament would tend to stay in the area of the filament. This forms a 'space charge' or a cloud of electrons whose mutual repulsion resists any further emission from the filament. The anode, being at ground potential, is more positive than the filament and attracts the electrons away from the filament - providing the primary acceleration for the electron beam. But the current flow that would result would be very low and dependent on the accelerating voltage. By adding the grid, or Wehnelt, we have a way of controlling the space charge of the filament, shaping the beam and increasing the beam current.

Fig. 2 - electrostatic potentials in SEM electron gun

Figure 2 depicts the equipotential force lines between the various parts of an electron gun. The equipotential lines form a contour map of the intensity, or flux, of the electrical field. The gradient, or direction, of the electrical field will always be perpendicular to these lines and is the direction of the most rapid change of the potential. Electrons leaving the filament will be accelerated along the gradient towards the most positive area, the anode. This beam of electrons will be focussed by the shape of the field gradient to a cross-over just before the anode, forming the first optical image of the source and ensuring that a larger percentage of the electrons will pass through the aperture of the anode.

Adjusting the bias resistance, and thereby the voltage differential between the filament and the grid, allows the beam current to be adjusted from a small de-focussed beam current, through a focussed maximum current, to cut-off. Cut-off is that point at which the more strongly negative fields of the grid prevent any electrons from reaching the anode by reversing the gradient completely around the filament.

While the plain tungsten wire filament is the most common cathode material in use, there are several variations and different materials used. Oxide and thoriated coatings have been explored to increase the emissivity of tungsten. Such coatings have not found much commercial use.

2.02 LaB6 cathode

In 1951, Lafferty2 established that the rare earths, and particularily lanthanum hexaboride (LaB6), had high thermionic emission characteristics and sufficiently low vapor pressures to be desirable cathode materials for electron microscopy.

LaB6 cathodes are becoming quite common. They use a single crystal LaB6 rod, of approximately 1 mm in diameter. This cathode can not be heated as directly as a tungsten filament, so a special mount or separate heater is used to provide the 1700 - 2100 degrees Kelvin required. The tip of the rod is polished to a point, then a small angled flat is usually polished at the point. The flat provides a defined area for emission. Without the flat, or if the cathode material evaporates past the flat, emission occurs from a broad undefined area around the point and resolution is decreased.

LaB6 cathodes provide around an order of magnitude higher brightness than tungsten cathodes. Longer cathode life is also an advantage, but both come at a cost. The LaB6 cathode requires a higher vacuum to operate (10-6 to 10-7 Torr) than the tungsten (10-5 Torr). This cleaner vacuum requirement can only be attained with different and more expensive vacuum system design and components.

A LaB6 electron gun uses the same basic cathode / grid / anode configuration used in a tungsten wire gun. Because of the high reactivity of the borides at the cathode temperatures used and their brittle nature, the mounting and heating of the cathode is crucial. Three different methods are in common use.

The Broers3,4,5 design utilizes a tungsten coil wrapped around the pointed end of a long (around 2 cm) LaB6 rod. This design uses the heat radiation and electron bombardment from the tungsten coil to heat the very end of the tip. The conduction of heat through the cathode holder, located at the other end of the rod, helps to lessen the problems of the reactivity of the material.

Vogel6 proposed a short LaB6 rod heated directly by passing a current through the LaB6 rod, perpendicular to the length of the rod. This is accomplished by using rigid electrical connectors that also provide the support for the rod. Pyrolytic graphite is used in between the conductors and the rod to avoid the problems of chemical reactivity with the LaB6 and the conductors.

In the design of Ferris et al.7, a short LaB6 rod is supported by a ribbon or strip through which an electrical current is passed for heating. The rod is heated by conduction from the ribbon. The ribbon material is chosen to be chemically inactive with the LaB6, such as graphite or tantalum.

Fig. 3 - LaB6 cathode construction

2.03 Field emission gun

The basic mechanism of field emission was discovered in 1897 by Wood8, who found that a high voltage applied between a pointed cathode and a plate anode caused a current to flow. Later, in 1937, Müller9 found that when a phosphor screen is placed some distance from a pointed high voltage source, a magnified image of the tip surface was formed.

These two discoveries illustrate two important properties. The first, that field emission results from electrons 'tunneling' past the work function of the metal tip helped by the high electrical field gradients. The second is that while electrons are emitted from the surface, their apparent source is a single point beneath the surface. That is, because of the electrical fields present, electrons tend to be emitted tangential to the surface which, in a hemispherical tip, results in an apparent source at the focus of the hemisphere. This apparent source will actually not be a point because of thermally caused differences in the electron's momentum.

Hibi10 first suggested in 1954 that a heated tungsten point, rather than a bent tungsten wire, might produce a smaller source size and higher brightness. This cathode actually incorporates both thermionic and field emissions and is referred to as a 'Schottky' or T-F (thermal - field) cathode.

Also in 1954, Cosslett and Haine11 proposed the use of a field emission cathode for electron microscopy. Due to the requirement for an extremely high vacuum, on the order of 10-9 Torr, no practical use was made until 1966. At that time, Crewe12, 13, 14 managed to build a usable system.

At the present time, field emission guns are gaining favor in electron microscopy as the costs of the associated very high vacuum equipment are going down.

Fig. 3 - Field emission electron gun

Fig. 3 shows a simplified schematic of a field emission (FE) gun. The FE tip is generally made of a single crystal tungsten wire sharpened by electrolytic etching. A tip diameter of 100 to 1000 Ċ is used, with the apparent source size much less than that. A simple tungsten tip can be very sensitive to surface contamination. More than any other cathode design, the field emission tip is extemely sensitive to the size, shape and surface condition. The electrostatic anodes are also very suseptable to contamination.

The emission process itself depends on the work function of the metal, which can be affected by adsorbed gases. This is the reason a very high vacuum is required.

Sustaining high electrical field gradients is also essential to emission, so a tip that is well worn might not emit electrons at all.

As with the tungsten filament gun, the voltage difference or bias between the first anode and the accelerating voltage on the cathode determines the emission current. The second anode is at ground potential and the voltage difference from here to the cathode determines the acceleration given the electrons. The shape of the anodes is carefully selected to minimize aberrations.

2.04 Cathode comparison

The following table gives simplified comparisons of the various cathodes in use.

In general, the smaller and brighter the apparent source, the higher the ultimate resolution. Many other factors are involved, in most cases a well designed modern instrument is limited more by the instrumental aberrations. In practical use, resolution of most instruments is limited by environmental effects - vibrations and electromagnetic fields.

Brightness of a cathode is defined as the current density per solid angle. More specifically:13

Equ. 1 - Gun brightness

Vacuum required relates primarily to the cost and complexity of an instrument. While a tungsten filament cathode can be used with a simple diffusion pump, the higher vacuum requirements can mean more exotic vacuum system pumps and seals.

SEM Cathode Comparison

Tungsten filament LaB6 Schottky (TF) Field Emission
Apparent Source Size 100 micrometers 5 micrometers <100 Angstroms <100 Angstroms
Brightness 1 A/cm2 steradian 20-50 A/cm2 steradian 100-500 A/cm2 steradian 100-1000 A/cm2 steradian
Vacuum Required 10-5 Torr 10-6 Torr 10-8 Torr 10-9 Torr
Table 1 - Gun comparisons


2.1 Electron beam lenses and apertures

Any electrical or magnetic field that affects a beam of electrons constitutes a lens. In this respect, the gun, deflection and stigmator coils of a scanning electron microscope are lenses, but in this section we will consider only the condensor and objective lenses. Electron optics are very similar in design and use to light optics. We'll start here with a description of light optics which might be more familiar to most.

In this discussion, we won't be attempting to relate an electron microscope to an optical microscope since no direct relationship can be made. Rather, we'll study the problem of producing an optical system capable of producing a small, focussed spot of light. This is directly analogous to the actions of the electron optics of an SEM, minus the beam deflection mechanism.

2.11 Thin lens light optics

In any optical system that starts and ends in the same medium, whether a single lens or multiple lenses, there are two planes perpendicular to the optical axis that are termed the principal planes. The location of these geometric constructs define one plane near the front of the lens system and one near the back where the location of these planes and the focal length of the system, measured from the planes, enable a complete characterization of the system.

The simplifications of the thin lens calculations come from the ability in many cases to consider the two principal planes to be coincident, or located at the same place. These calculations become more complex in the so called 'thick' lens, particularily in multiple lens systems. In this light optics section, we'll be using only thin lens examples.

To begin with, we must consider the source of the light. In the process of creating as small a final spot as possible, the apparent size of the source will determine the optics required. As an example, a source size of 1mm2 will require a magnification in the optics of 0.1x (the magnification of less than one is used here to represent the negative magnification, or de-magnification) in order to produce a final spot size of 0.1mm2. At the point of the final spot, an optical image of the source will be formed.

The term apparent size is used here to indicate that there can be many variables that affect the optical appearance of the source. A 1mm wide by 2mm long helical tungsten filament, for example, may appear as a 1mm by 1mm source of visible light when heated. This could be due to the conductive cooling of the outer portions of the coil by the structures that support the coil, thus preventing the emission of visible light from those areas. In this case, the physical size of the source is twice as large as the apparent size.

Fig. 4 - Real vs. virtual images

The resulting optical image we want to form is a real image. A real image is one that is formed outside of the optics system - essentially a projected image. A virtual image is one that is not projected outside of a system, but is rather an extension of the optical path back into the system. In the course of our discussions on light and electron optics, we will restrict ourselves to real images. The example in Fig. 4a above depicts the bending, or refraction, of light from a source back to a real image projected in front of the lens. This is a simplified representation of the focussing of a light source into a spot.

Refraction is based on the fact that light travels at different rates in different materials. The ratio of the speed of light in a vacuum to that in a particular material is known as the absolute refractive index: n=c/v where n is the refractive index, c the speed of light in a vacuum, and v the speed of light in the lens medium. A refractive index can also be calculated for any two materials, and for practical reasons is often calculated for a lens operated in air. The refractive index varies for a given material based on the wavelength of the light involved, the ratio being larger for blue light than red. However, the speed of light in a vacuum is constant for all wavelengths.

Fig. 5 - Refraction and refractive index

Fig. 5 demonstrates the basics of calculating light refraction. For computational simplification, light is normally considered to consist of a rays traveling in the same direction as the light. A - A' represents a plane wave of light traveling from the left and being refracted at the interface with medium n'. The refraction, or change in direction of the incoming light, is demonstrated here as the difference between the lengths of segments A - B and A' - B'. Because light moves at a slower rate in medium n', the segment A' - B' is shorter than the corresponding segment A - B, where this same time interval is spent in medium n. This results in the change of direction from A - A' to B - B'. The angle I is the incident angle of the impinging wave and angle I' is the angle of refraction.

From this example it is clear that the index of refraction and the incident angle are the determining factors in the refraction of light. The incident angle will affect the length of segment A - B and the ratio of the index of refraction for each material will affect the length of segment A' - B'. From this we can derive n'/n=AB/A'B'. By dividing the numerators and denominators by BA', we can bring this formula to trigonometric terms as n'sin I' = nsin I (Snell's Law).

Fig. 6 - Representative lens parameters

In this simplified discussion of optics, we'll only be considering symetrical lenses and using thin lens methods. Thin lens calculations are first order approximations valid only when the thickness of the lens is negligible compared to it's focal length. Fig. 6 shows several parameters we'll be discussing. F is the Effective Focal Length, or the point at which a group of parallel rays coming into the lens from the left would converge from the refraction of the lens. This is normally just refered to as the focal length, as it will be here.

In Fig. 6 the distances U and V are shown equal to each other, and in this case the image magnification would be 1. Magnification is related to these distances by: M=V/U where M is the magnification. Some other ways of determining magnification are:

M=V/U
M=(V-F)/F
M=F/(U-F)

The distances, U and V, and the overall distance D can be determined from:

U=V/M
U=(R*D)/(R+1)
U=D/(M+1)
U=(F*V)/(V-F)

V=U*M
V=U/R
V=(F/R)+F
V=(F*M)+F

D=F*(M+(1/M)+2)
D=(F*(M+1)2)/M
D=F*(R+(1/R)+2)
D=(F*(R+12)/R

The focal point can also be determined from the parameters shown by:

F=(U*M)/(M+1)
F=(D*R)/(R+1)2
F=(D*M)/(M+1)2
F=U/(R+1)
F=(U*V)/D
F=1/((1/U)+(1/V))

From these equations, we can determine the operating characteristics of simple optical systems. For example, let's say we have a need for a single lens system that will provide the smallest image magnification if the distance D from the object to image is 100mm and the lens can be no closer than 5mm to the image (V). It's clear from M=V/U that putting the lens at the minimum 5mm from the image would give the smallest magnification, in this case 5/95 or a magnification of 0.053.

In order for us to use this setup, and have the image focussed, we'll need to select a lens with an appropriate focal length. Using F=(U*M)/(M+1) we find that a lens with a focal length of 4.78mm would fit.

Obviously, it's not possible to design a single lens system, given these parameters, that would produce a smaller image size. But what if we used more than one lens? For simplicity, let's split the 100mm length we have into two 50mm sections. We'll put one lens into each section, in each case 45mm from the light entrance of each section. Now, using M=V/U we get a magnification of 0.11 for each section.

Fig. 7 - Two lens system

In this setup, the first section produces a reduced image at the start of the second section, which then becomes the source for the next section. To calculate the total magnification of this new system, we merely have to multiply the magnifications of the two systems together, which gives us a magnification of 0.0121, which would produce an image that's 1/5 the size of the the single lens solution. Using the same formula to determine the focal length for the lenses needed would give us a focal length of 4.46mm.

From the examples above, you can see how we're able to produce a greatly demagnified image of a light source through the use of a multi-lens system. This system could be extended to even more lenses, but there are other considerations that affect any optical design and can place severe constraints, particularily on multiple lens systems. The common definition for aberration is, according to Webster's, "deviation from the normal or usual. In optics, however, aberrations are quite the norm. The term aberration in optics refers to a number of effects that prevent light rays being refracted from acting ideally. Specifically, aberrations are classified into spherical, coma, astigmatism, curvature of field and distortion. Aberrations can be inherent in the design and materials used and they can be caused by imperfections in manufacture. Aberrations will be covered in more detail later.

2.12 Electron optics

There are a number of differences between light and electron optics. As we've seen, light optics relies on the refraction (or reflection) of light. In electron optical systems, we use electrostatic or magnetic fields to influence the tragectories of beams of electrons. There is no definite interface in the refracting medium of fields as there is in the solid lenses of light optics. Additionally, electrons in the influence of a magnetic field will also rotate around the optical axis, an effect not seen in light optics. Moving electrons in a vacuum will travel in straight paths as does light, but where there are a number of electrons moving together, their electrical mutual repulsion will cause the beam to diverge.

In practical instruments used today, electrostatic lenses are used only in the electron gun, as described in the previous section. Magnetic lenses are used through the rest of an instrument. Electrostatic lenses require conducting surfaces very close to the path of an electron beam in order to produce an electrical field of high intensity. These surfaces must be accurately formed, extremely smooth and are easily contaminated. Magnetic fields, on the other hand, are usually formed by solenoid coils that are located completely outside of the vacuum system. Therefore, they suffer none of the contamination problems inherent in electrostatic lenses. Electrostatic lenses, however, can be made extremely small and are capable of producing much faster response for beam deflections. For these reasons, they are also used for beam blanking, a technique very useful in the SEM analysis of integrated circuits.


Fig. 8 - Electrostatic lens (adapted from Feynman, et.al.15)

A high electrical field gradient will accelerate a charged particle such as an electron. In a scanning electron microscope, that fact is used in the electron gun assembly to provide the initial acceleration to the beam electrons. Electrical fields can also be used for focussing a beam of charged particles. In Fig. 8 above, a series of three cylinders with their axis coincident with the beam axis, demonstrate the principal of electrostatic focussing (please note that the cylindrical shape of the electrodes is merely representative - they could be the inner surfaces of apertures in disc shaped electrodes). Electrons entering into the system from the left will be affected by the electrical field formed by the large voltage difference between the first two cylinders. The electrons will be given both an impulse towards the cylinder axis and a boost in velocity by the increasingly negative field. As the electrons move into the second cylinder, they receive an impulse towards the walls of the cylinder, but since they are now closer to the axis and traveling faster, the change in direction will be less than from the first impulse.

Entering the second field, the electrons will once again be given an impulse towards the cylinder walls. As they pass through the field, though, they will be de-accelerated by the increasingly positive field. Finally, they will again receive an impulse towards the cylinder axis. When exiting the system, the electron beam will again have it's original velocity, and there will be a net impulse towards the cylinder axis, resulting in a focussing of the beam. If the center cylinder were at a positive voltage in relation to the other two, it would create a diverging lens system.

Magnetic fields can also be used as lenses for beams of charged particles. In an SEM, electro-magnetic lens are used almost exclusively for the condensor and objective lenses (some instruments refer to the objective lens as the final condensor lens, but I prefer using the term objective for this lens which is usually of a considerably different design than the condensors). Electro-magnetic lenses used in commercial SEMs have always been solenoid coils. While so-called 'strong focussing' or multiple pole lenses are theoretically possible, they have not been used primarily due to the increased complexity of accurately manufacturing them.

A solenoid is a cylindrical coil with circumferencially wound wire. In SEMs, the lens is completely encased in a ferromagnetic shroud forming a thick walled cylinder, with a small gap around the inner circumference. The gap confines the external magnetic field to a small area and may be located anywhere on the inner surface. The strength of the magnetic field generated is proportional to the number of turns of wire multiplied by the current passing through the coil (N*I where N = number of wire turns, I = current). Relatively large fields are required for focussing electron beams of the accelerating voltages found in SEMs (up to 50KV).


Fig. 9 - Electro-magnetic lens

Fig. 9 represents a number of features of a magnetic lens. Magnetic field lines are shown, as are equipotential lines. These give a visible description of the shape of the magnetic field. The light blue lines represent two possible electron paths. The smaller inset gives an approximation of the magnetic field strength along the optical axis and perdendicular to the optical axis. What can not be faithfully represented in a two dimensional image is the rotation of charged particles about the axis in a uniform magnetic field. This effect is the heart of the cyclic particle accelerators, momentum spectrometer and other instrumental techniques.

In a uniform magnetic flux density B, a moving charged particle will follow a curved path . For a point charge, like an electron, the force imparted on it by electrical and magnetic fields is found in the vector equation F = q(E+v*B) (the Lorentz force) where F is the force, q is the charge on the particle, E is the electric field, v the particle's velocity and B is the magnetic field. In the case of a magnetic lens, there is no electric field and the E term is dropped giving F = qv*B. As an electron comes down the optical path, it first encounters the horizontal component (Hr)of the magnetic field. This causes the electron to begin a rotation about the axis. The strengthening of the vertical field component (HZ) gives the electron an impulse towards the axis. Finally, as the electron emerges through the bottom horizontal field component, it receives an impulse that stops it's rotation about the axis. Note that unlike the case of electrostatic lenses, no change in velocity takes place along the optical axis.


Fig. 10 - Imaging in the SEM

2.3 Aberrations


References, Section 2

1. P. D. Zemany, In Symp. X-ray Electron Probe Analy. ASTM Spec. Tech. Publ. 349 (1964)

2. J. M. Lafferty, J. Appl. Phys. 22, 299-309 (1951)

3. A. N. Broers, J. Appl. Phys. 38, 1991-1992 (1967)

4. A. N. Broers, U. S. Patent 3,462,635, Aug. 19, 1969

5. A. N. Broers, J. Phys. E. 2, 273-276 (1969)

6. S. F. Vogel, Rev. Sci. Inst. 41, 585-587 (1970)

7. S. D. Ferris, D. C. Joy, H. J. Leamy and C. K. Crawford, IITRI/SEM 1975, 12-18 (1975)

8. R. W. Wood, Phys. Rev., 5, (1897)

9. E. W. Müller, Z. Physik, 106, 541-550 (1937)

10. T. Hibi, Proc. Int. Conference on Electron Microscopy, London, 1954, Royal Microscopical Society, London, 636-638 (1956)

11. V. E. Cosslett and M. E. Haine, Proc. Int. Conference on Electron Microscopy, London, 1954, Royal Microscopical Society, London, 639-644 (1956)

12. A. V. Crewe, Science, 154, 729-738 (1966)

13. A. V. Crewe, D. N. Eggenberger, J. Wall and L. M. Welter, Rev. Sci. Instr., 39, 576-583 (1968)

14. A. V. Crewe and J. Wall, J. Mol. Biol., 48, 373-393 (1970)

15. R. P. Feynman, R. B. Leighton and M. Sands, The Feynman Lectures on Physics, Vol. 2, 29-4 (1964)